![]() |
|
HealthEast Care System Chemical Indicator Project
EXPERIMENTATION EQUIPMENT: Steris-Amsco® Eagle 3000 60” Deep Dynamic Steam Sterilizer
Determine most appropriate or most challenging locations inside various
Synthes® Instrument Set cases, Major Instrument and Bookwalter Retractor
sets for the placement of internal chemical indicators. The purpose of
this is to minimize the chance that these sets would be released when
the sterilant (saturated steam) did not penetrate the most challenging
areas.
Created some failures, but were not able to determine precisely where the most challenging location was for placing the chemical indicators. Where the medical device manufacturer’s instruction is not available, a single chemical indicator in the center of the set is acceptable. Only the medical device manufacturer can determine through validation
testing (spore seeding, thermocouple use) where the most challenging areas
are for placement of chemical indicators.
Reinstate chemical indicator monitoring program for instrument sets,
packs and load. Meet best practice standards as established by the Association
for the Advancement of Medical Instrumentation (AAMI), for accreditation
reasons (JCAHO) and in the event of a legally required need situation.
Implement change at the acute care hospitals, ambulatory surgery centers
and office-based physician clinics.
Revise appropriate system policies. Develop system protocols for type
and placement in set, package and load, retrieval and interpretation and
action in the event of a non-responsive result. To be accomplished in
the Fall 2005.
Michael Festa Director of Sterile Processing HealthEast Care System mfesta@healtheast.org 651.232.3106
|
||||||||
Copyright © MHCSMA 2003, All Rights Reserved. |